Metrel MI 6601 MediTest Medical Electrical Equipment Safety Tester
- Brand
- Metrel
- Model number
- MI 6601
- Category
- Electrical Safety Testers
Medical device safety tester to IEC/EN 62353 and IEC 60601: protective earth continuity, insulation, equipment, applied part and patient leakage currents, power and harmonics; built-in test socket and printer support.
Price and availability on request
Key features
- PE resistance with 200 mAAC and 25 AAC
- Insulation resistance with test voltage 250 VDC or 500 VDC
- All test configurations acc. to IEC 62353 are supported
- Leakage current measurements with 1 µA resolution
- AC, DC and TRMS value of leakage currents
- All leakage current measurements as defined in IEC 60601-1 standard (patient, auxiliary, earth, touch)
- All leakage current measurements as defined in IEC 62353 (equipment and applied part; alternative, direct, differential method)
- Measurement according to portable appliance standards EN 50678 and EN 50699
- Point-to-point, touch and mains voltage tests
- Equipment power
- IEC lead test
- Pre-set or configurable visual and functional inspections
- The most complete IEC 60601 tester on the market
- 10 universal configurable connections. They can be configured as applied parts, non-earthed parts, earthed parts
- Dedicated test ports for connection to functional earthing and signal I/O connections
- All possible configurations acc. to the IEC 60601 and IEC 62353 standards
- Tests can be run on the instrument directly or via a PC
- Fully automatic test flow with Auto Sequences®
- Tests and limits are automatically set acc. to the set configuration of medical equipment
- Test sequences are optimized for fastest work flow
- Supports measuring by standards IEC/EN 60601 (2nd and 3rd editions), IEC/EN 62353, ANS/NZS 3551, EN 50678 and EN 50669
- Use of single tests to diagnose problems easily
Specifications
| Protective earth resistance | 0.00 Ω ... 999 Ω; ±(2 % of reading + 2 D) up to 19.99 Ω |
|---|
| Insulation resistance (250 V, 500 V) | 0.00 MΩ ... 199.9 MΩ; ±(3 % of reading + 2 D) up to 19.99 MΩ |
|---|---|
| Insulation LN-PE, LN-NEP, LN-AP, AP-PE, AP-NEP | 0.00 MΩ ... 199.9 MΩ |
| Substitute leakage current | 0.00 mA ... 19.99 mA; ±(3 % of reading + 3 D) up to 1.99 mA |
|---|---|
| Differential, PE, touch and equipment leakage | 0.000 mA ... 19.99 mA; ±(3 % of reading + 3 D) up to 1.999 mA |
| Active / apparent power | 0 W ... 3.70 kW / 0 VA ... 3.70 kVA |
|---|---|
| THDU / current | 0.0 % ... 99.9 % / 0.00 A ... 16.00 A |
| Clamp current | 0.10 mA ... 24.9 A |
|---|
| Voltage | 0.0 V ... 264 V |
|---|
| Dimensions (w×d×h) / weight | 42 cm x 33 cm x 18 cm / 8.1 kg |
|---|
| MI 6601 Standard set | Instrument MI 6601 MediTest; 1x A 1080 Power cable, 2m, 3x1.5mm2; Test lead, Black, 1m, 1.5mm2 (A 1758); Test lead, Brown, 1m, 1.5mm2 (A 1759); Test lead, Green, 1m, 1.5mm2 (A 1760); Test lead, Yelow, 1m, 1.5mm2 (A 1761); Test lead, Violet, 1m, 1.5mm2 (A 1762); Test probe, Black (A 1014); Test probe, Brown (A 1298); Test probe, Green (A 1062); 2x Crocodile clip, Black (A 1013); Crocodile clip, Brown (A 1297); Crocodile clip, Green (A 1309); Crocodile clip, Yellow (A 1546); USB cable L=1m (A 1727 |
|---|
Specifications are from the manufacturer's published information. Confirm critical values with us or the datasheet before ordering.