Metrel MI 6601 MediTest Medical Electrical Equipment Safety Tester

Brand
Metrel
Model number
MI 6601

Medical device safety tester to IEC/EN 62353 and IEC 60601: protective earth continuity, insulation, equipment, applied part and patient leakage currents, power and harmonics; built-in test socket and printer support.

Price and availability on request

Request a quote

Sent with the product details: Metrel MI 6601 MediTest Medical Electrical Equipment Safety Tester.

A phone number or an email address is needed, so we can reply.

We use these details only to reply to your enquiry. See our privacy notice.

Key features

  • PE resistance with 200 mAAC and 25 AAC
  • Insulation resistance with test voltage 250 VDC or 500 VDC
  • All test configurations acc. to IEC 62353 are supported
  • Leakage current measurements with 1 µA resolution
  • AC, DC and TRMS value of leakage currents
  • All leakage current measurements as defined in IEC 60601-1 standard (patient, auxiliary, earth, touch)
  • All leakage current measurements as defined in IEC 62353 (equipment and applied part; alternative, direct, differential method)
  • Measurement according to portable appliance standards EN 50678 and EN 50699
  • Point-to-point, touch and mains voltage tests
  • Equipment power
  • IEC lead test
  • Pre-set or configurable visual and functional inspections
  • The most complete IEC 60601 tester on the market
  • 10 universal configurable connections. They can be configured as applied parts, non-earthed parts, earthed parts
  • Dedicated test ports for connection to functional earthing and signal I/O connections
  • All possible configurations acc. to the IEC 60601 and IEC 62353 standards
  • Tests can be run on the instrument directly or via a PC
  • Fully automatic test flow with Auto Sequences®
  • Tests and limits are automatically set acc. to the set configuration of medical equipment
  • Test sequences are optimized for fastest work flow
  • Supports measuring by standards IEC/EN 60601 (2nd and 3rd editions), IEC/EN 62353, ANS/NZS 3551, EN 50678 and EN 50669
  • Use of single tests to diagnose problems easily

Specifications

Continuity
Protective earth resistance0.00 Ω ... 999 Ω; ±⁠(2 % of reading + 2 D) up to 19.99 Ω
Insulation
Insulation resistance (250 V, 500 V)0.00 MΩ ... 199.9 MΩ; ±⁠(3 % of reading + 2 D) up to 19.99 MΩ
Insulation LN-PE, LN-NEP, LN-AP, AP-PE, AP-NEP0.00 MΩ ... 199.9 MΩ
Leakage
Substitute leakage current0.00 mA ... 19.99 mA; ±⁠(3 % of reading + 3 D) up to 1.99 mA
Differential, PE, touch and equipment leakage0.000 mA ... 19.99 mA; ±⁠(3 % of reading + 3 D) up to 1.999 mA
Power
Active / apparent power0 W ... 3.70 kW / 0 VA ... 3.70 kVA
THDU / current0.0 % ... 99.9 % / 0.00 A ... 16.00 A
Current
Clamp current0.10 mA ... 24.9 A
Voltage
Voltage0.0 V ... 264 V
General
Dimensions (w×d×h) / weight42 cm x 33 cm x 18 cm / 8.1 kg
Supplied with
MI 6601 Standard setInstrument MI 6601 MediTest; 1x A 1080 Power cable, 2m, 3x1.5mm2; Test lead, Black, 1m, 1.5mm2 (A 1758); Test lead, Brown, 1m, 1.5mm2 (A 1759); Test lead, Green, 1m, 1.5mm2 (A 1760); Test lead, Yelow, 1m, 1.5mm2 (A 1761); Test lead, Violet, 1m, 1.5mm2 (A 1762); Test probe, Black (A 1014); Test probe, Brown (A 1298); Test probe, Green (A 1062); 2x Crocodile clip, Black (A 1013); Crocodile clip, Brown (A 1297); Crocodile clip, Green (A 1309); Crocodile clip, Yellow (A 1546); USB cable L=1m (A 1727

Specifications are from the manufacturer's published information. Confirm critical values with us or the datasheet before ordering.