Masibus TC12+ tCAL Temperature Calibrator
- Brand
- Masibus
- Model number
- TC12+
- Category
- Process Calibrators
Masibus TC12+ tCAL temperature calibrator. For laboratory and site calibration purpose, measure and simulate for thermocouple.
Price and availability on request
Key features
- Compact, handheld, user friendly menu
- Easy to read color graphical TFT LCD display
- Rechargeable lithium ion battery with enhanced power control for prolonged battery life
- Measure: mA/ V/ mV/ mA (24V)/ Switch-test / RTD/ TC
- Source: Resistance/ RTD/ TC
- 24 VDC Loop power supply to power transmitters and loops
- Step/Ramp functions with auto/ man selection
- Universal Serial Bus (USB) communication port for charging, data retrieve and firmware upgrade
- Data logging to measure long time drift
- Other Features: Max/ Min/ Average, filter settings, tare facility, adjustable backlight, alarm annunciation (on display and buzzer), glance screen mode
- Continuity test
- Pulsed RTD transmitter compatible
- HART loop resistor
Specifications
| Resistance (Ohms) | 0 to 400 Ω; 400 to 4000Ω # |
|---|---|
| Pt10 to Pt1000 | -200 to 200 °C; 200 to 600 °C; 600 to 850 °C |
| Ni100 | -60 to 180 °C |
| Ni120 | -80 to 260 °C |
| V | 0 to 30.00 VDC / 0.001 V |
| mA | 0 to 24.000 mA / 0.001 mA Thermocouple/mV Measurement/Simulation Resolution & Accuracy@20-30°C |
| TC Type | Range / Resolution |
| E | -200.0 to 1000.0 °C / 0.1 °C |
| J | -200.0 to 1200.0 °C / 0.1 °C |
| K | -200.0 to 1372.0 °C / 0.1 °C |
| T | -200.0 to 400.0 °C / 0.1 °C |
| B | 450.0 to 1800.0 °C / 0.1 °C |
| R | 0.0 to 1750.0 °C / 0.1 °C |
| S | 0 to 1750.0 °C / 0.1 °C |
| N | -200.0 to 1300.0°C / 0.1 °C |
| mV | -10.000 to 80.000 m / V 0.001 mV; -10.00 to 250.00 m V 0.01mV Note: Temperature standard ITS-90 Degree equivalent to 4uV against respective readings to be added to above mentioned accuracy for TC input. |
| Display Mode | Measure: mA/ V/ mV/ mA(2W)/ Switch-test / RTD/ TC Source: Resistance/ RTD/ TC |
|---|---|
| Supported units for RTD/ TC type | °C/ °F/ °K |
| RTD Measurement Current | 300 uA |
| Settling time (pulsed currents RTD Simulation) | >1 ms |
| CJC error (For Thermocouple) Internal Reference Junction | ≤± 0.5 °C |
| CJC selection | Manual/ Internal/ External(1) |
| Max. input voltage (EM Terminal) | 30 VDC |
| Temperature Coefficient | ≤30 ppm |
| Input Impedance | TC/ mV/ V >1MΩ mA: 10 Ω |
| Response time | Input <100ms, Output <100ms |
| Load impedance | >4.7KΩ for TC/mV |
| Display update rate | 10 readings / sec |
| Isolation | 500VDC between mA/V Measure and RTD /Ω /TC/mV |
| Data logging | Logged data is stored in a user defined file in internal memory Periodic logging: 150000 readings max |
| Communication Interface | USB 2.0 (1)with RTD sensor at RTD terminal for External CJC |
| Pt10 to Pt400 0.01°C Pt500, Pt1000 0.1°C | 4 wire Measurement: ±0.15°C, Simulation*: ±0.15 °C; 4 wire Measurement: ±0.2 °C, Simulation*: ±0.25 °C; 4 wire Measurement: ±0.3 °C, Simulation*: ±0.35 °C |
|---|
| Applications | Laboratory and site calibration purpose, Measure and simulate for thermocouple, Calibration of Transmitters and Transducers, Drift test of Transmitters and Transducers |
|---|---|
| Industries | Calibration Lab, Process plant, Dairy plant, Sensor manufacturer (TC/RTD), Mining and Materials industries, Engineering laboratories, Electronics Industries |
Specifications are from the manufacturer's published information. Confirm critical values with us or the datasheet before ordering.