| Circuit | Custom one-chip of microprocessor LSI; circuit. |
|---|
| Display | LCD size : 52 mm x 38 mm; LCD with green backlight ( ON/OFF ). |
|---|
| Measurement Range | 30 - 130 dB. |
|---|
| Resolution | 0.1dB. |
|---|
| Function | dB ( A & C frequency weighting ); Time weighting ( Fast, Slow ); Peak hold, Data hold.; Record ( Max., Min. ). |
|---|
| Accuracy | Characteristics of " A " frequency weighting network meet IEC 61672-1-2013 class 1; Under 94 dB input signal, the accuracy are: 31.5Hz reading± 1.5 dB; 63Hz reading± 1.0 dB; 125Hz reading± 1.0 dB; 250Hz reading± 1.0 dB; 500Hz reading± 1.0 dB; 1KHz reading± 0.7 dB; 2KHz reading± 1.0 dB; 4KHz reading± 1.0 dB; 8KHz reading+ 1.5 dB -2.5dB; 12.5KHz reading+ 2.0 dB -5.0dB; 16KHz reading+ 2.5 dB -16.0dB; Remark : The above spec. are tested under the; environment RF Field Strength less |
|---|
| Accuracy (cont.) | than 3 V/M & frequency less than 30; MHz only. |
|---|
| Frequency Weighting Network | Characteristics of A & C.; A weighting : The characteristic is simulated as "Human Ear Listening" response. Typical, if making the environmental sound level measurement, always select to A weighting.; C weighting; The characteristic is near the "FLAT" response. Typical, it is suitable for checking the noise of machinery (Q.C. check) & knowing the sound pressure level of the tested equipment. |
|---|
| Data hold | Freeze the display reading. |
|---|
| Peak hold | To keep the peak ( max. ) measurement value. |
|---|
| Time weighting | Fast - t= 125 ms; * "Fast" range is simulated the human ear response time weighting.; Slow - t = 1 Sec.; * "Slow" range is easy to get the average values of vibration sound level. |
|---|
| Range selector | Auto range : 30 to 130 dB.; Manual range : 3 range, 30 to 80 dB, 50 to 100 dB, 80 to 130 dB, 50 dB on each step, with over & under range indicating. |
|---|
| Frequency | 31.5 to 16 KHz. |
|---|
| Microphone type | Electric condenser microphone. |
|---|
| Microphone size | Out size, 12.7 mm DIA. ( 1/2 inch). |
|---|
| Calibration VR | Build in external calibration VR, easy to calibrate on 94 dB level by screw driver.; * Calibrated via external SOUND CALIBRATOR ( SC-941, optional ). |
|---|
| Calibrator | B & K (Bruel & kjaer), Multifunction Acoustic Calibrator 4226. |
|---|
| Datalogger Sampling Time Setting range | Auto 1,2,5,10,30,60,120,300,600, 1600,3800seconds #If I set to 1 second,memory data may loss. |
|---|
| Memory Card | SD memory card. 1 GB to 16 GB. |
|---|
| Advanced setting | * Set clock time ( Year/Month/Date,Hour/Minute/ Second ); * Decimal point of SD card setting; * Auto power OFF management; * Set beep Sound ON/OFF; * Set sampling time; * SD memory card Format |
|---|
| Over Indication | Show " - - - - ". |
|---|
| Memory Recall | Maximum & Minimum value. |
|---|
| Sampling Time of Display | Approx. 1 second. |
|---|
| Data Output | RS 232/USB PC computer interface.; * Connect the optional RS232 cable; UPCB-02 will get the RS232 plug.; * Connect the optional USB cable; USB-01 will get the USB plug. |
|---|
| Power off | Auto shut off (approx. 10 minutes ) saves battery life or manual off by push button. |
|---|
| Operating Temperature | 0 to 50 °C. |
|---|
| Operating Humidity | Less than 85% R.H. |
|---|
| Power Supply | * Alkaline or heavy duty DC 1.5V battery.; ( UM3, AA ) x 6 PCs, or equivalent.; * DC 9V adapter input. ( AC/DC power; adapter is optional ). |
|---|
| Power Current | Normal operation ( w/o SD card save data and LCD Backlight is OFF) : Approx. DC mA.; When SD card save the data but and LCD Backlight is OFF) : Approx. DC 3 mA.; * AC/DC power adapter is optional. consumption will increase approx. 8mA |
|---|
| Weight | 320 g/0.70 LBS.(without battery) |
|---|
| Dimension | 245 x 68 x 45 mm.; ( 9.6 x 2.7x 1.8 inch ). |