Fluke 1577 Insulation Multimeter
- Brand
- Fluke
- Model number
- 1577
- Category
- Insulation Multimeters
Insulation tester (500 V, 1000 V, 0.1 to 600 MΩ) and true-rms multimeter in one handheld unit.
Price and availability on request
Key features
- Insulation test voltages 500 V, 1000 V
- Insulation test: 0.1 MΩ to 600 MΩ
- Auto discharge of capacitive voltage
- AC/DC Voltage and milliamps
- Three-year warranty
Specifications
| Voltage DC, Maximum voltage | 1000 V |
|---|---|
| Voltage DC, Accuracy | ±(0.2%+2) |
| Voltage DC, Maximum resolution | 0.001 V |
| Voltage AC, Maximum voltage | 1000 V |
| Voltage AC, Accuracy | ±(2%+3) |
| Voltage AC, Maximum resolution | 0.1 mV |
| Current DC, Maximum amps | 400 mA |
| Current DC, Amps accuracy | ±(1.0%+2) |
| Current DC, Maximum resolution | 0.01 mA |
| Current AC, Maximum amps | 400 mA |
| Current AC, Amps accuracy | ±(2.0%+2)* |
| Current AC, Maximum resolution | 0.01 mA |
| Resistance, Maximum resistance | 50 MΩ |
| Resistance, Accuracy | ±(1.2%+2) |
| Resistance, maximum resolution | 0.1Ω |
| Minimum test current @ 1kΩ/V | 1 mA |
|---|---|
| Test voltages | 500, 1000 V |
| Maximum resistance per test voltage | 50 V; 100 V; 250 V; 500 V; 1000 V; 600 MΩ |
| Maximum resolution per test voltage | 50 V; 0.01 MΩ; 100 V; 0.01 MΩ; 250 V; 0.1 MΩ; 500 V; 0.1 MΩ; 1000 V; 0.1 MΩ |
| Accuracy per test voltage | 50 V; ±(3% + 5); 100 V; ±(3% + 5); 250 V; ±(1.5% + 5); 500 V; ±(2.0% + 5); 1000 V; ±(2.0% + 5) |
| Operating temperature | –20°C to +55°C |
|---|---|
| Storage temperature | -40°C to +60°C |
| Humidity (without condensation) | 0% - 95% (10°C – 30°C); 0% - 75% (30°C – 40°C); 0% - 40% (40°C – 55°C) |
| Operating altitude | 2000 m |
| General | IEC 61010-1: Pollution Degree 2 |
|---|---|
| Measurement | IEC 61010-2-033: CAT IV 600V; CAT III 1000V |
| Size | 203 x 100 x 50 mm (with holster) |
|---|---|
| Weight | 624 g |
| Warranty | 3 years |
| Battery life alkaline, Meter use | 1000 hours |
| Battery life alkaline, Insulation test use | Meter can perform at least 1000 insulation tests with fresh alkaline batteries at room temperature. These are standard tests of 1000 V into 1 MΩ with a duty cycle of 5 seconds on and 25 seconds off. |
| IP rating | IP40 |
| Contents | Remote probe; Test leads; Alligator clips; Hard case; Four AA alkaline batteries; Quick reference guide; CD rom with manual |
|---|
Specifications are from the manufacturer's published information. Confirm critical values with us or the datasheet before ordering.
About the 1577
The Fluke 1577 combines a digital insulation tester and a true-rms multimeter in one compact unit for motors, generators, cables and switchgear. It tests insulation at 500 V and 1000 V from 0.1 MΩ to 600 MΩ, discharges capacitive voltage automatically and measures AC/DC voltage and milliamps.