HTC Instruments DIM-1504 Digital Insulation Multimeter

Model number
DIM-1504

Insulation tester and 5,999-count true RMS multimeter in one: test voltages 100 V to 2500 V, insulation resistance up to 50 GΩ, polarization index and absorption ratio.

Price and availability on request

Request a quote

Sent with the product details: HTC Instruments DIM-1504 Digital Insulation Multimeter.

A phone number or an email address is needed, so we can reply.

We use these details only to reply to your enquiry. See our privacy notice.

Key features

  • 5,999 counts
  • True RMS
  • Insulation resistance test at 100 V, 250 V, 500 V, 1000 V and 2500 V
  • Polarization index (PI)
  • Absorption ratio
  • Auto range switching
  • Non-contact voltage (NCV) detection
  • Duty cycle
  • Data hold
  • Dual integrating A/D converter
  • Auto power off
  • Backlight

Specifications

Insulation resistance
Test voltages100 V, 250 V, 500 V, 1000 V, 2500 V
Measurement range100 V: 0.1 MΩ – 2 GΩ; 250 V: 0.25 MΩ – 5 GΩ; 500 V: 0.5 MΩ – 10 GΩ; 1000 V: 1 MΩ – 20 GΩ; 2500 V: 2.5 MΩ – 50 GΩ
Accuracy100 V: ±⁠(3% + 5); other test voltages: 0.25 – 399.9 MΩ ±⁠(3% + 5); 0.4 MΩ – 2 GΩ ±⁠(5% + 5); 2.001 – 10 GΩ ±⁠(8% + 5); 10.01 – 50 GΩ ±⁠(10% + 5)
Display ranges (auto)100 V: 4 M / 40 M / 400 M / 2 GΩ; 250 V: up to 5 GΩ; 500 V: up to 10 GΩ; 1000 V: up to 20 GΩ; 2500 V: 40 M / 400 M / 4 G / 40 G / 50 GΩ
Resolution0.001 MΩ – 0.001 GΩ (2500 V: 0.01 MΩ – 0.01 GΩ)
Short-circuit current≥ 1 mA
Open-circuit voltage90 – 110% of rated voltage
Drop-off resistor / alarm100 V: 0.2 MΩ (alarm < 0.2 MΩ); 250 V: 0.5 MΩ (< 0.5 MΩ); 500 V: 1 MΩ (< 1 MΩ); 1000 V: 2 MΩ (< 2 MΩ); 2500 V: alarm < 5 MΩ
DC voltage
Range (resolution)6 V – 1000 V (0.001 V – 1 V)
Accuracy±⁠(0.5% + 3)
Input impedanceApprox. 10 MΩ
Maximum input1000 V
AC voltage (true RMS)
Range (resolution)6 V – 750 V (0.001 V – 1 V)
Accuracy±⁠(0.8% + 5)
Frequency range40 Hz – 1 kHz
Input impedanceApprox. 10 MΩ
Maximum input750 V
DisplayTrue RMS (sine-wave RMS calibration)
Resistance
Ranges (resolution)600 Ω – 6 MΩ (0.1 Ω – 0.001 MΩ); 60 MΩ (0.01 MΩ)
Accuracy600 Ω – 6 MΩ: ±⁠(0.5% + 3); 60 MΩ: ±⁠(2% + 5)
Overload protection250 V RMS
Capacitance
Ranges (resolution)1 – 10 nF (0.001 – 0.01 nF); 100 nF – 1 mF (0.1 nF – 1 µF); 10 – 100 mF (0.01 – 0.1 mF)
Accuracy1 – 10 nF: ±⁠(3% + 20); 100 nF – 1 mF: ±⁠(3% + 5); 10 – 100 mF: ±⁠(8% + 5)
Overload protection250 V RMS
Diode and continuity
DiodeDisplays approximate forward voltage; forward current approx. 1 mA
ContinuityBuzzer sounds below approx. 50 Ω; open-circuit voltage approx. 1 V
Overload protection250 V DC or AC peak
General
Safety categoryCAT III 600 V
Power6 × 1.5 V AA
DimensionsH 200 × W 90 × D 50 mm
WeightApprox. 480 g
AccessoriesTest leads, protective holster, alligator clip test lead, manual and batteries

Specifications are from the manufacturer's published information. Confirm critical values with us or the datasheet before ordering.