Delta IMA-S2000-48-YV 48V 41.6A 2000W Enclosed Power Supply, IMA Series
- Brand
- Delta
- Model number
- IMA-S2000-48-YV
IMA enclosed power supply with 48V 41.6A output (2000W) from 80 - 275 Vac or 120 - 300 Vdc. Fan cooling, operating -20 to +70 °C. With medical safety design and current sharing.
Price and availability on request
Key features
- Safety approval for Medical (IEC 60601-1-2 and 3rd+A1 edition CB report) TUV EN60601-1:2006
- CAN/CSA-C22.2 NO.60601-1 + ANSI/AAMI, ES 60601-1, 3000m altidude, 100-240Vac only
- Industrial (IEC 62368) and IT IEC60950-1 CB report, IEC62368-1 CB report, TUV EN62368-1, UL62368-1 and CAN/CSA C22.2 No. 62368-1
- 3000 Watt peak power (up to 5s)
- Standby power < 2 Watt
- Wide operating input voltage range: 80 Vac to 275 Vac or 120 Vdc to 300 Vdc
- Wide adjustable output voltage range (+/- 20%)
- Standby output 5 Vdc / 2 A
- High efficiency: up to 94%
- Size: 309.6 x 127 x 40.6 mm (12.2 x 5 x 1.6 in) (1U design)
- Low acoustic noise level of less than 39 dB(A), variable speed fan control
- Active current sharing
- 2 × MOPP and BF appliances
- PMBus™ compatible for control, programming and monitoring
- 500,000 hour MTBF
- Optional conformal coating
- 3 years warranty
Specifications
| Series | IMA Series |
|---|---|
| Mounting | Enclosed |
| Output Current | 41.6A |
|---|---|
| Output Voltage | 48V |
| Output Voltage Range | 48Vdc |
| Output Voltage Adjustment Range | 38.4 V – 57.6 V |
| Output Power | 2000W |
| Hold-up Time | 20 ms |
| Start-up with Capacitive Loads | 10,000 μF |
| Residual Ripple/PARD (20MHz) | < 480 mV |
| Rise Time | < 200 ms |
| Nominal Input Voltage | 100 - 240 Vac |
|---|---|
| Input Voltage Range | 80 - 275 Vac |
| Nominal Input Frequency | 50 - 60 Hz |
| Input Frequency Range | 47-63Hz |
| DC Input Voltage Range | 120 - 300 Vdc |
| Input Current | 17.2 A @ 80 Vac / 12 A @ 120 Vdc |
| Inrush Current | < 25 A |
| Power Factor at 100% load | > 0.97 (typical) |
| Leakage Current | AC Leakage current from Input to earth ground / AC Leakage current from Output to earth ground |
| Indicator | Green LED |
|---|---|
| Cooling System | 2 fans with variable speed control |
| Input Terminal | Block M4 x 3 pins |
| Output Terminal | Block M5 x 2 pins |
| Signal Terminal | Connector x 14 pins |
| Noise Sound Pressure Level (SPL) | < 39 dB(A) |
| Unit Weight | 2 kg (4.41 lb) |
| Dimensions | 309.6 x 127 x 40.6 mm / 12.19” x 5.00” x 1.60” |
| Operating Temperature | -20 °C to +70 °C (-4°F to +158 °F) |
|---|---|
| Storage Temperature | -40 °C to +85 °C (-40 °F to +185 °F) |
| Power De-rating | When AC input voltage is < 110 Vac, the output power is reduced by 20 W per 1 V. |
| Shock Test (Non-Operating) | IEC 60068-2-27 compliant, 50 g, 11 ms, 3 shocks for each direction |
| Vibration (Non-Operating) | IEC 60068-2-6 compliant, 2.09 Grms, 5 - 500 Hz, 20 minutes per side (3 planes) |
| Operating Altitude | 5000 m (54kPa) |
| Pollution Degree | 2 |
| Overvoltage | YES, latch mode / 58 V to 61 V |
|---|---|
| Overload / Overcurrent | 3000 W for over 750 ms max / Standby 3.2 A max with Hiccup Mode (Non-Latching, Auto-Recovery) |
| Over Temperature | YES, auto recovery |
| Short Circuit | YES, auto recovery |
| MTBF | 500,000 hours |
|---|---|
| Expected Cap Life Time | 10 years |
| Safety Certificate | IEC 62368 Edition 2: / IEC 62368-1 (2014) Edition 2 / 5000 m (16,400 ft) altitude, 120 V to 300 Vdc and 100 V to 240±10% Vac / / IEC/EN 60950-1, Edition 2 and all national deviations: / UL 60950-1/CSA 22.2 No 60950-1, Edition 2; UL 62368-1 / 5000 m (16,400 ft) altitude, 120 V to 300 Vdc and 100 V to 240 ±10% Vac (UL File E135494) / / IEC/EN 60601-1, Edition 3: / IEC 60601-1(2005), EN60601-1(2006) / ANSI/AAMI ES 60601-1(2005) / CAN/CSA C22.2 No. 60601-1 (2008); 3,000 m (9,800 ft) altitude, 100 V |
|---|
| Emission | EN 55032, EN 55032 and FCC, Class B |
|---|---|
| Harmonic Current Emissions | EN 61000-3-2, Class A |
| Voltage Fluctuation and Flicker | EN 61000-3-3 |
| ESD Immunity | EN 61000-4-2, level 4, 8 kV contact, 15 kV air |
| Radiated Immunity | EN 61000-4-3, level 3, 12 V/m |
| EFT | EN 61000-4-4, level 4, ±2 kV |
| Surge | EN 61000-4-5, level 4, 1 kV DM, 2 kV CM |
| Conducted Immunity | EN 61000-4-6, level 3, 10 Vrms |
| Magnetic Field | EN 61000-4-8, level 3, 10 A/m |
| Voltage Dips | EN 61000-4-11, 30%, 20 ms - R&D / EN 61000-4-11, 60%, 100 ms / EN 61000-4-11, 100%, 500 ms / EN 60601-1-2:2015 (4th edition), 30%, 500 ms / EN 60601-1-2:2015 (4th edition), 60%, 100 ms / EN 60601-1-2:2015 (4th edition), 100%, 20 ms / EN 60601-1-2:2015 (4th edition), 100%, 5000 ms |
Specifications are from the manufacturer's published information. Confirm critical values with us or the datasheet before ordering.