Masibus TT7S10-XP Loop Powered Temperature Transmitter
- Brand
- Masibus
- Model number
- TT7S10-XP
- Category
- Sensors and Transmitters
Masibus TT7S10-XP loop powered temperature transmitter: output ±0.25% of FS, 13.5-36 V DC, 2 Wire supply. For industrial process control, factory automation.
Price and availability on request
Key features
- Universal input (RTD, Thermocouple, Ohm, mV)
- Available in wall or 2″ Pipe mount Ex-proof Transmitter with touch sense keys0.56″, 4 digit LED display in Single Compartment0.39″, 4 digit LED display in Dual Compartment
- 1.5 KV RMS Isolation between Input & Output
- Linearized Output
- Highly Accurate
- Fully Programmable for Input type & Range
- Digital Filter
- Easy configuration using keys and display
- Windows mTRAN software for Configuration, Calibration & Monitoring
- Reverse Loop polarity protection
- Direct / Reverse output
- Sensor break detection
Specifications
| RTD | Pt100, 3/4-Wire |
|---|---|
| Resistance/Potentiometer | 0-2500Ω |
| Sensor Current | 0.2 mA |
| Thermocouple | E,J,K,T,B,R,S,N with Internal CJC (ANSI Standard) |
| Linear | 0-75 mV/0-500 mV DC/ 4-20mA/0-20mA |
| Input Impedance for V Input | > 1MΩ |
| Input Burden for mA Input | ≤ 1V |
| Sensor Break current | < 1 uA |
| Input Range | Refer Table -1 |
| Zero / Span Adjust | Adjustable either from Keyboard or from mTRAN Software |
| E,J,K,T,N,Pt100 | 0.1% of FS, ±1 °C |
|---|---|
| B,R,S | 0.25% of FS, ±1 °C |
| Linear, Potentiometer | 0.1% of FS, ±1 Unit |
| E,J,K,T,N | ±2 °C |
|---|---|
| B,R,S | ±3 °C |
| ADC Resolution | 16 bits |
| Stability | ± 0.1% per year |
| Response time | < 1 sec approx., if filter is set to 0 |
| Digital Filter | 0-20 settable through software (2 default) |
| Sensor open Detection | Available (Except 0-20mA I/P) |
| Allowable wiring resistance for RTD | Maximum 15 ohms/wire (Resistance between all wires should be equal) |
| CMRR | >120 dB |
| NMRR | ~40 dB |
| Temp-co | <150 ppm |
| Display size | 0.56" / 0.39" |
|---|---|
| Display type | 7 segment, Red LED, 4 digits |
| Keys | 3 keys (ENT, ESC, INC) for configuration, calibration and operation |
| Current Output (2 wire) | 4-20mA or 20-4mA (User programmable, Scaled by OPLO and OPHI) |
|---|---|
| Output accuracy | ±0.25% of FS |
| Sensor break Output | <3.8 or >20.8mA programmable |
| Output load | R load = (V supply-13.5)/0.021Ω |
| Interface | TTL (3wire), Protocol-Modbus RTU |
|---|---|
| Baud Rate | 9600 bps |
| Power Supply | 13.5-36 V DC, 2 Wire |
|---|
| Galvanic Isolation | 1500V AC for 1 minute between input & output |
|---|
| Mounting | Wall (Std) or 2" Pipe mount (optional) |
|---|---|
| Dimensions in mm | 140(H) x 145(W) x 80(D) / 100(H) x 100(W) x 145(D) |
| Weight (without mounting clamps) | 1 Kg / 1.65 Kg |
| Enclosure Material | Aluminum Alloy LM-6 |
| Ingress Protection | IP65 / IP66 |
| Area Classification | Zone 1 & 2, Gas Group: IIA & IIB |
| Cable entry size | 2 nos of M20 double compression cable glands |
| Storage temperature | -20 to 85°C Humidity 30% to 95% RH(Non-condensing) Table-1: Display Range |
|---|---|
| RTD | Pt100 3/4 wire / -200 to 850.0°C |
| Linear | 0-75mV DC 0-500mV DC -1999 to 99 4-20 mA 0-20 mA; 0-20 mA |
| Resistance/ Potentiometer | 0-2500Ω; -1999 to 99 |
| Applications | Industrial process control, Factory automation, Interface SCADA, Heat treatment furnaces, Reheat furnaces, Ceramic Kilns, Glass Industry, Water and waste water control |
|---|---|
| Industries | Power Plants, Metal Industry, Oil & Gas, Chemical, Glass Industry, Cement, Fertilizer, Paper, Pulp |
Specifications are from the manufacturer's published information. Confirm critical values with us or the datasheet before ordering.