Rishabh AECT 12 Smart Earth Clamp Tester
- Brand
- Rishabh
- Model number
- AECT 12
- Category
- Earth Resistance Testers
Clamp-on earth resistance tester for looped grounding systems: measures 0.01 Ω to 1500 Ω without auxiliary electrodes or disconnecting the earth lead; 65 x 32 mm jaw, 99-reading memory.
Price and availability on request
Key features
- No auxiliary electrodes and no disconnection needed in grounding systems with a loop
- Earth resistance up to 1500 Ω
- Internal data logging up to 99 readings
- Wide 65 x 32 mm jaw for conductors and earthing strips
- Pass / fail alarm with sound and light (limit 1-199 Ω)
- Applications: multiple earth pits (mesh / grid), earth bonding, loose or broken earthing strips, telecom tower grounding
Specifications
| Ranges | 0.010-0.099 Ω (0.001 Ω) ±(1% + 0.1 Ω); 0.10-0.99 Ω ±(1% + 0.1 Ω); 1.0-49.9 Ω ±(1% + 0.5 Ω); 50.0-99.5 Ω ±(1.5% + 0.5 Ω); 100-199 Ω ±(2% + 1 Ω); 200-395 Ω ±(5% + 5 Ω); 400-590 Ω ±(10% + 10 Ω); 600-880 Ω ±(20% + 20 Ω); 900-1500 Ω ±(25% + 30 Ω) |
|---|
| Single measurement time | 0.5 s |
|---|---|
| Measurement frequency | > 1 kHz |
| Working conditions | -20 °C to 55 °C, 10% - 90% RH |
| Dimensions | 285 mm (L) x 85 mm (W) x 56 mm (H) |
| Weight | 1100 g including battery |
| Jaw size | 65 mm x 32 mm |
|---|
| LCD | 4 digit, 47 mm x 28.5 mm |
|---|
| Supply | 6 V DC (4 alkaline cells) |
|---|
| Protection | Double insulation, CAT III 150 V, CE |
|---|
| Contents | Standard loop, instruction manual, aluminium case |
|---|
Specifications are from the manufacturer's published information. Confirm critical values with us or the datasheet before ordering.