Megger MIT410/2 CAT IV Insulation and Continuity Tester
- Brand
- Megger
- Model number
- MIT410/2
- Category
- Insulation Resistance Testers
Handheld 1 kV insulation tester with five test voltages from 50 V to 1000 V, readings up to 200 GΩ, PI and DAR, fast 0.01 Ω to 1 MΩ continuity and 600 V True RMS voltage; CAT IV 600 V, IP54.
Price and availability on request
Key features
- Insulation testing at 50 V, 100 V, 250 V, 500 V and 1000 V, up to 200 GΩ at 1000 V
- Stabilised insulation test voltage, maintained within -0% +2% (industry standard 10-20%)
- Polarisation Index (PI, 10 min / 1 min) and Dielectric Absorption Ratio (DAR, 60 s / 15 s or 60 s / 30 s)
- Timed testing and insulation test lock
- PASS / FAIL indication on limit alarms
- Leakage current display
- Single auto-ranging continuity range from 0.01 Ω to 1 MΩ, with 200 mA or 20 mA test current and automatic bi-directional testing
- True RMS AC and DC voltage measurement to 600 V with automatic frequency display
- Analogue arc plus dual digital display (result and test voltage, leakage current, frequency or test mode)
- Live circuit detection inhibits insulation testing above 25 V, 30 V, 50 V, 75 V or 100 V (default 50 V), even if the fuse has failed
- Over-moulded case sealed to IP54, including battery and fuse compartment
- Runs on 6 AA alkaline or NiMH rechargeable cells
Specifications
| Test voltages | 50 V, 100 V, 250 V, 500 V, 1000 V |
|---|---|
| Range and accuracy | 10 GΩ at 50 V, 20 GΩ at 100 V, 50 GΩ at 250 V, 100 GΩ at 500 V, 200 GΩ at 1000 V; ±2% ±2 digits plus 4.0% (50 V), 2.0% (100 V), 0.8% (250 V), 0.4% (500 V) or 0.2% (1000 V) per GΩ |
| Resolution | 0.1 kΩ |
| Short circuit / charge current | 2 mA +0% -50% to EN 61557-2 (2007) |
| Open circuit voltage | -0% +2% ±2 V |
| Test current | 1 mA at min. pass value of insulation, 2 mA max. |
| Leakage current | 10% ±3 digits |
| Timer | 60 second countdown timer |
| Analogue display | 1 GΩ full scale |
| Range | 0.01 Ω to 1000 kΩ |
|---|---|
| Accuracy | ±3% ±2 digits (0 to 100 Ω), ±5% ±2 digits (100 Ω to 500 kΩ) |
| Test current | 200 mA (-0 mA +20 mA) from 0.01 Ω to 4 Ω, or 20 mA |
| Open circuit voltage | 5 V ±1 V |
| Lead null | Up to 9.00 Ω |
| Buzzer limit | Adjustable from 1 Ω to 200 Ω in 12 steps |
| Range | 0.0 mV to 600 V AC and DC, True RMS |
|---|---|
| Accuracy | DC ±2% ±2 digits; AC ±2% ±2 digits (>10 V TRMS sinusoidal, 15 Hz to 400 Hz) |
| Frequency | 15 Hz to 400 Hz, ±0.5% ±1 digit (15 Hz to 99.9 Hz) |
| Input impedance | 0.25 MΩ |
| Power | 6 x 1.5 V AA (IEC LR6) alkaline cells; 6 x 1.2 V NiMH rechargeable cells may be used |
|---|---|
| Battery life | 3000 insulation tests (5 s on / 55 s off at 1000 V into 1 MΩ) |
| Fuse | 1 x 500 mA (FF) 1000 V 32 x 6 mm ceramic HBC fuse, 30 kA minimum |
| Safety | EN 61010-1, 600 V phase to earth, CAT IV; EN 61557 |
| IP rating | IP54 |
| Operating temperature | -10 °C to +55 °C, 90% RH at 40 °C max. |
| Storage temperature | -25 °C to +70 °C |
| Maximum altitude | 2000 m |
| Dimensions | 228 x 108 x 63 mm |
| Weight | 1.75 kg (instrument and case) |
| Order code | 1004-735 |
| Contents | Red / black silicone test leads with probes and clips; SP5 remote switch probe; 6 x AA alkaline batteries; Hard carry case; Calibration certificate |
|---|
Specifications are from the manufacturer's published information. Confirm critical values with us or the datasheet before ordering.