Lutron YK-22DOA Dissolved Oxygen Meter

Brand
Lutron
Model number
YK-22DOA

Dissolved oxygen meter for 0 to 20.0 mg/L and oxygen in air, with a polarographic probe on a 4 m cable, ATC and height compensation.

Price and availability on request

Request a quote

Sent with the product details: Lutron YK-22DOA Dissolved Oxygen Meter.

A phone number or an email address is needed, so we can reply.

We use these details only to reply to your enquiry. See our privacy notice.

Key features

  • One meter for Dissolved Oxygen measurement.
  • Dissolved oxygen: 0 to 20.0 mg/L.
  • Dissolved oxygen meter use the polar graphic type oxygen probe with temperature sensor, high precision measurement for Dissolved Oxygen (DO) and temperature measurement.
  • Heavy duty dissolved oxygen probe, probe head can connect with BOD bottle.
  • DO use the automatic Temp. compensation.
  • DO use the automatic Temp. compensation.
  • DO meter build in " % SALT " & " Mountain Height " compensation value adjustment.
  • Separate probe, easy for operation of different measurement environment.
  • Wide applications: water conditioning, aquariums, beverage, fish hatcheries, food processing, pHotograpHy, laboratory, paper industry, plating industry, quality control, school & college, water conditioning.
  • LCD with green light backlight, easy reading.
  • Can default auto power off or manual power off.
  • Data hold, record max. and min. reading.
  • Microcomputer circuit, high accuracy.
  • Power by DC 9V battery (006 P) or DC 9V adapter.
  • RS232/USB PC COMPUTER interface.

Specifications

Specifications
DisplayLCD size : 52 mm x 38 mm; LCD with green backlight ( ON/OFF ).
CircuitCustom one-chip of microprocessor LSI circuit.
Measurement FunctionDissolved Oxygen, Temperature.
Advanced setting* Auto power OFF management; * Set beep Sound ON/OFF; * Set temperature unit to °C or °F; * Set DO salt % compensation value; * Set DO height ( meter ) compensation value; * Set DO height ( feet ) compensation value
Data HoldFreeze the display reading.
Memory RecallMaximum & Minimum value.
Sampling Time of DisplayApprox. 1 second.
Data OutputRS 232/USB PC computer interface.; * Connect the optional RS232 cable UPCB-02 will get the RS232 plug.; * Connect the optional USB cable USB-01 will get the USB plug.
Operating Temperature0 to 50 °C.
Operating HumidityLess than 85% R.H.
Power Supply* Alkaline or heavy duty DC 9V ( 006 P ) battery or equivalent.; * DC 9V adapter input. ( AC/DC power adapter is optional ).
Power CurrentNormal operation ( LCD Backlight is OFF) : Approx. DC 14 mA.; * If LCD backlight on, the power consumption will increase approx. 12 mA.
Weight489 g/1.08 LB.
Dimension177 x 68 x 45 mm (7.0 x 2.7x 1.9 inch)
Probe Weight335 g/0.74 LB
Probe Size190 mm x 28 mm Dia. ( 7.5" x 1.1" Dia. )
Probe Cable Length4 meters.
Accessories
Supplied with* Operation manual : 1 PC; * Hard carrying case, CA-06 : 1 PC; * Dissolved Oxygen probe, OXPB-11 : 1 PC; *Spare Probe head with Diaphragm OXHD-04 : 2 PCs; *Probe-filling Electrolyte OXEL-03 : 1 PC
Optional* Oxygen probe : OXPB-11; * Spare Probe head with Diaphragm set : OXHD-04; * Probe-filling Electrolyte : OXEL-03; * AC to DC 9V adapter.; * USB cable, USB-01.; * RS232 cable, UPCB-02.; * Data Acquisition software, SW-E816.; * Excel Data Acquisition software, SW-E802.
Measurement Ranges
Dissolved Oxygen0 to 20.0 mg/L
Oxygen in Air0 to 100.0%
Resolutions
Dissolved Oxygen0.1 mg/L
Oxygen in Air0.1%
Temperature0.1°C
Accuracy
Dissolved Oxygen±0.4 mg/L
Oxygen in Air±0.7%
Temperature±0.8°C / ±1.5°F
Technical details
Height CompensationAdjustable from 0 to 8,900 meters

Specifications are from the manufacturer's published information. Confirm critical values with us or the datasheet before ordering.