Lutron BTM-4208SD 12-channel Temperature Recorder, SD Card

Brand
Lutron
Model number
BTM-4208SD
Category
Data Loggers

12-channel thermocouple temperature recorder for type J/K/T/E/R/S probes, logging all channels with time stamps to an SD card.

Price and availability on request

Request a quote

Sent with the product details: Lutron BTM-4208SD 12-channel Temperature Recorder, SD Card.

A phone number or an email address is needed, so we can reply.

We use these details only to reply to your enquiry. See our privacy notice.

Key features

  • 12 channels Temperature recorder, use SD card to save the data along with time information, paperless.
  • Real time data logger, save the 12 channels Temp. measuring data along the time information (year, month, date, minute, second) into the SD memory card and can be down load to the Excel, extra software is no need. User can make the further data or graphic analysis by themselves.
  • Channels no.: 12 channels (CH1 to CH12) temperature measurement.
  • Sensor type: Type J/K/T/E/R/S thermocouple.
  • Auto datalogger or manual datalogger. Data logger sampling time range: 1 to 3600 seconds.
  • Type K thermometer: -100 to 1300 °C.
  • Type J thermometer: -100 to 1200 °C.
  • Page select, show CH1 to CH8 or CH9 to CH12 in the same LCD.
  • Display resolution: 1 degree/0.1 degree.
  • Offset adjustment.
  • SD card capacity: 1 GB to 16 GB.
  • RS232/USB computer interface.
  • Microcomputer circuit provides intelligent function and high accuracy.
  • Jumbo LCD with green light backlight, easy reading.
  • Can default auto power off or manual power off.
  • Data hold to freeze the measurement value.
  • Record function to present the max. and min. reading.
  • Power by UM3/AA (1.5 V) x 8 batteries or DC 9V adapter.
  • RS232/USB PC COMPUTER interface.
  • Heavy duty & compact housing case.

Specifications

Specifications
CircuitCustom one-chip of microprocessor LSI circuit.
DisplayLCD size : 82 mm x 61 mm.; * with green color backlight.
Channels12 channels : T1, T2, T3, T4, T5, T6, T7, T8, T9, T10, T11 and T12.
Sensor type ResolutionType K thermocouple probe.; Type J/T/E/R/S thermocouple probe.; 0.1 °C/1 °C, 0.1 °F/1°F.
Datalogger Sampling Time Setting rangeAuto: 1 second to 3600 seconds; @ Sampling time can set to 1 second, but memory data may loss.; Manual : Push the data logger button once will save data one time.; @ Set the sampling time to 0 second.
Memory CardSD memory card. 1 GB to 16 GB.
Advanced setting* Set clock time ( Year/Month/Date, Hour/Minute/ Second ); * Decimal point of SD card setting; * Auto power OFF management; * Set beep Sound ON/OFF; * Set temperature unit to °C or °F; * Set sampling time; * SD memory card Format
Temperature CompensationAutomatic temp. compensation for the type K/J/T/E/R/S thermometer.
Linear CompensationLinear Compensation for the full range.
Offset AdjustmentTo adjust the zero temperature deviation value.
Probe Input Socket2 pin thermocouple socket.
Over IndicationShow " - - - - ".
Data HoldFreeze the display reading.
Memory RecallMaximum & Minimum value.
Sampling Time of DisplayApprox. 1 second.
Data OutputRS 232/USB PC computer interface.; * Connect the optional RS232 cable UPCB-02 will get the RS232 plug.; * Connect the optional USB cable USB-01 will get the USB plug.
Power offAuto shut off saves battery life or manual off by push button.
Operating Temperature0 to 50 °C.
Operating HumidityLess than 85% R.H.
Power Supply* Alkaline or heavy duty DC 1.5 V battery ( UM3, AA ) x 8 PCs, or equivalent.; * DC 9V adapter input. ( AC/DC power adapter is optional ).
Power CurrentNormal operation ( w/o SD card save data and LCD Backlight is OFF) :Approx. DC 7.5 mA.; When SD card save the data but and LCD Backlight is OFF) : Approx. DC 25 mA.; * If LCD backlight on, the power consumption will increase approx. 11 mA.
WeightMeter : 948g ( includes batteries )
Dimension225 X 125 X 64 mm ( 8.86 X 4.92 X 2.52 inch )
Accessories
Supplied with* Instruction manual:1 PC; * Type K Temp. probe, TP-01:2 PC; * SD Card ( 2 GB ):1 PC; * Hard carrying case, CA-08:1 PC
Optional* Type K thermocouple probe. TP-01, TP-02A. TP-03, TP-04; * USB cable, USB-01.; * RS232 cable, UPCB-02; * Data Acquisition software, SW-U811-WIN.; * EXCEL data Acquisition software, SW-E802.; * AC to DC 9V adapter.
Measurement Functions
Type K-100 to 1300°C
Type J-100 to 1200°C
Type T-100 to 400°C
Type E-100 to 900°C
Type R0 to 1700°C
Type S0 to 1500°C
Accuracy±⁠(0.4% + 1°C) for most ranges

Specifications are from the manufacturer's published information. Confirm critical values with us or the datasheet before ordering.