| DC Voltage range | 60.000 mV to 1000.0 V, 6 ranges, Basic accuracy: ±0.025% rdg. ±2 dgt. |
|---|
| AC Voltage range | 60.000 mV to 1000.0 V, 6 ranges, Frequency characteristics: 20 Hz - 100 kHz Basic accuracy 45 - 65 Hz : ±0.2% rdg. ±25 dgt. (True RMS, crest factor 3) |
|---|
| DC + AC Voltage range | 6.0000 V to 1000.0 V, 4 ranges, Frequency characteristics: 20 Hz - 100 kHz Basic accuracy 45 - 65 Hz : ±0.3% rdg. ±30 dgt. (True RMS, crest factor 3) |
|---|
| Resistance range | 60.000 Ω to 600.0 MΩ, 8 ranges, (Conductance: 600.00 nS, DT4282 only) Basic accuracy: ±0.03% rdg. ±2 dgt. |
|---|
| DC Current range | 600.00 μA to 10.000 A, 6 ranges, Basic accuracy: ±0.05% rdg. ±5 dgt. |
|---|
| AC Current range | 600.00 μA to 10.000 A, 6 ranges, Basic accuracy 45 - 65 Hz : ±0.6% rdg. ±5 dgt. (True RMS, crest factor 3) Frequency characteristics: 20 Hz - 20 kHz (at 600 μA to 600 mA range) |
|---|
| AC Current range (use with Clamp on probes) | N/A |
|---|
| Peak | DC V measurement: Signal width 4 msec or more (single), 1 msec or more (repeated) AC V, DC/AC A measurement: Signal width 1 msec or more (single), 250 μsec or more (repeated) |
|---|
| Capacitance range | 1.000 nF to 100.0 mF, 9 ranges, Basic accuracy: ±1.0% rdg. ±5 dgt. |
|---|
| Continuity Check | Continuity threshold: 20/50/100/500 Ω, Response time: 10 ms or more |
|---|
| Diode test | Open terminal voltage: 4.5 V or less, Testing current 1.2 mA or less, Threshold of forward voltage: 0.15 V to 3 V, seven stages |
|---|
| Frequency range | AC V, DC+AC V, AC A measurement, at pulse width 1 μs or more (50% duty ratio) 99.999 Hz (0.5 Hz or more) to 500.00 kHz, 5 ranges, ±0.005% rdg. ±3 dgt. |
|---|
| dB conversion | Standard impedance setting (dBm), 4 Ω to 1200 Ω, 20 stages Display dB conversion value of AC voltage (dBV) |
|---|
| Temperature (thermocouples) | K: -40.0°C to 800.0°C (-40.0°F to 1472.0°F) Add accuracy of the Thermocouple probe to main unit accuracy: ±0.5% rdg. ±3°C |
|---|
| Other functions | Filter function (remove harmonic noise, use only at 600 V AC, 1000 V AC ranges), display value hold, auto hold, MAX/MIN value display, PEAK value display, relative value display, sampling setting, internal memory (400 data), auto-power save, USB communication (option), mis-insertion prevention shutters, decibel conversion, 4-20 mA percentage conversion |
|---|
| Display | Main and Sub displays: 5-digits LCD, max. 60000 digits |
|---|
| Display refresh rates | 5 times/s (Capacitance measurement: 0.05 to 2 times/s, depending on measured value, Temperature: 1 time/s ) |
|---|
| Power supply | LR6 (AA) alkaline batteries ×4, Continuous use: 100 hours |
|---|
| Dimensions and mass | 93 mm (3.66 in.) W × 197 mm (7.76 in.) H× 53 mm (2.09 in.) D, 650 g (22.9 oz.) (with test leads holder and batteries) |