| DC Voltage range | 600.0 mV to 1000 V 5 ranges, Basic accuracy: ±0.3% rdg. ±3 dgt. |
|---|
| AC Voltage range | 6.000 V to 1000 V, 4 ranges, Frequency characteristics: 40 Hz to 1 kHz Basic accuracy 40 - 500 Hz : ±0.9% rdg. ±3 dgt. (True RMS, crest factor 3) |
|---|
| AUTO AC/DCV | Yes |
|---|
| Resistance range | 600.0 Ω to 60.00 MΩ, 6 ranges, Basic accuracy: ±0.7% rdg. ±3 dgt. |
|---|
| DC Current range | N/A |
|---|
| From 4 to 20mA Percentage conversion display | N/A |
|---|
| AC Current range (use with Clamp on probes) | 10.00 A to 1000 A, 7 ranges, Add the Clamp on probe accuracy to basic accuracy 40 - 1 kHz : ±0.9% rdg. ±3 dgt. (True RMS, crest factor 3) |
|---|
| Temperature (thermocouples) | N/A |
|---|
| Voltage detection | Hi: 40 V to 600 V AC, Lo: 80 V to 600 V AC |
|---|
| Capacitance range | 1.000 μF to 10.00 mF, 5 ranges, Basic accuracy: ±1.9% rdg. ±5 dgt. |
|---|
| Frequency range | 99.99 Hz to 99.99 kHz, 4 ranges (limited by the minimum detectable voltage), Basic accuracy: ±0.1% rdg. ±1 dgt. |
|---|
| Continuity check | Continuity threshold [ON]: 25 Ω or less , [OFF]: 245 Ω or more, Response time: 0.5 ms or more |
|---|
| Diode test | Open terminal voltage: 5.0 V or less, Testing current 0.5 mA or less, Threshold of forward voltage: 0.15 V to 1.5 V |
|---|
| Other functions | Filter function, display value hold, auto hold, MAX/MIN/Average value display, relative value display, auto-power save, USB communication (option) |
|---|
| Display | Main and Sub displays: 4-digits LCD, max. 6000 digits, bar graph |
|---|
| Display refresh rates | 5 times/s (Capacitance measurement: 0.05 to 5 times/s, depending on measured value, Frequency: 1 to 2 time/s) |
|---|
| Power supply | LR03 alkaline batteries ×4, Continuous use: 130 hours (backlight OFF) |
|---|
| Dimensions and mass | 84 mm (3.31 in.)W × 174 mm (6.85 in.)H× 52 mm (2.05 in.)D, 390 g (13.8 oz.) (including batteries and holster) |