Hioki CM4373-50 AC/DC Clamp Meter

Brand
Hioki
Model number
CM4373-50

True RMS AC/DC clamp meter for 2000 A with stronger jaws and an IP54 dust and water resistant body; inrush current measurement and DC voltage to 2000 V with the P2010 probe. Z3210 wireless adapter sold separately.

Price and availability on request

Request a quote

Sent with the product details: Hioki CM4373-50 AC/DC Clamp Meter.

A phone number or an email address is needed, so we can reply.

We use these details only to reply to your enquiry. See our privacy notice.

Key features

  • CAT IV 600 V, CAT III 1000 V When using P2010: CAT III 2000 V, CAT IV 1000 V
  • Measure DC voltages of up to 2000 V for open voltage inspections of solar panels
  • Simultaneously measure inrush current in RMS and crest values
  • Easily transfer measurement data to your smartphone or tablet by using our free app GENNECT Cross or to an Excel® file
  • Harmonic analysis from 1st to 30th order with GENNECT Cross
  • Automatic AC/DC function helps boost work efficiency

Specifications

Specifications
DC Current range600.0 A/2000 A, Basic accuracy: ±1.3% rdg. ±0.3 A (600 A range)
AC Current range600.0 A/2000 A (10 Hz to 1 kHz, True RMS), Basic accuracy 45 - 66 Hz: ±1.3% rdg. ±0.3 A (at 600 A)
Crest factor600.0 A range: 3 or less 2000 A range: 2.84 or less
DC+AC Current range600.0 A/2000 A (10 Hz to 1 kHz, True RMS), Basic accuracy DC, 45-66 Hz: ±1.3% rdg. ±1.3 A (at 600 A)
DC Voltage range600.0 mV to 1000 V (When using P2010: 600.0 V to 2000 V)
AC Voltage range6.000 V to 1000 V, 4 ranges (15 Hz to 1 kHz, True RMS), Basic accuracy 45 - 66 Hz: ±0.9% rdg. ±0.003 V (at 6 V)
DC+AC Voltage range6.000 V to 1000 V, 4 ranges, Basic accuracy DC, 45 - 66 Hz: ±1.0% rdg. ±0.013 V (at 6 V)
Resistance range600.0 Ω to 6.000 MΩ, 5 ranges, Basic accuracy: ±0.7% rdg. ±0.5 Ω (at 600 Ω)
Capacitance range1.000 μF to 1000 μF, 4 ranges, Basic accuracy: ±1.9% rdg. ±0.005 μF (at 1 μF)
Frequency range9.999 Hz to 999.9 Hz, 3 ranges, Basic accuracy: ±0.1% rdg. ±0.003 Hz (at 9.999 Hz)
Temperature (K)-40.0 to 400.0 ˚C, add temperature probe accuracy to basic accuracy of ±0.5% rdg. ±3.0 ˚C
Voltage detectionHi: 40 V to 600 V AC, Lo: 80 V to 600 V AC, 50/60 Hz
Harmonic levels[With Z3210 installed] Voltage/Current harmonic levels up to 30th, Content factor, Total harmonic distortion ratio *Harmonics can be displayed with our free app GENNECT Cross.
Other functionsDC power, Continuity check, Diode check, Automatic AC/DC detection, Pass/fail judgement function of DC A and DC V, Max/Min/Average/PEAK MAX/ PEAK MIN value display, Low-pass filter function, Display value hold, Auto hold, Back light, Auto-power save, Buzzer sounds, Zero-adjustment, etc.
Dustproof, waterproofIP20 (Voltage measurement in a completely dry condition. When jaw closes) IP54 (While in storage)
Power supplyLR03 Alkaline battery ×2 Continuous use: 40 hr (without Z3210 installed), 24 hr. (with Z3210 installed and using wireless communications) Other conditions: 100 A AC measurement, backlight off, 23°C reference value
Core jaw diameterφ55 mm (2.17 in), Jaw dimension: 92 mm (3.62 in) W×18 mm (0.71 in) D
Dimensions and mass65 mm (2.56 in) W×250 mm (9.84 in) H×35 mm (1.38 in) D mm, 530 g (18.7 oz)
Supplied with
ContentsTest Lead L9300, Carrying Case C0203, LR03 Alkaline battery ×2, Instruction Manual×2, Operating Precautions ×1
Ordering
Order codeCM4373-50 (Wireless Adapter Z3210 not included)

Specifications are from the manufacturer's published information. Confirm critical values with us or the datasheet before ordering.