Fluke 707 Loop Calibrator with Quick Click Knob
- Brand
- Fluke
- Model number
- 707
- Category
- Process Calibrators
High-performance mA loop calibrator with quick-click knob for one-handed use; simultaneous mA and % readout.
Price and availability on request
Key features
- High performance loop calibrator for calibration, repair, and maintenance
- Provides simultaneous mA and % readout for quick, easy interpretation of readings
- Delivers mA accuracy of 0.015%
- Provides 1 µA resolution for mA source, simulate, and measure
Specifications
| Current measurement, Range | 0 to 24 mA |
|---|---|
| Current measurement, Resolution | 0.001 mA |
| Current measurement, Accuracy | 0.015% Rdg + 2 LSD |
| Current sourcing, Range | 0 to 20 mA or 4 to 20 mA |
| Current sourcing, Accuracy | 0.015% Rdg + 2 LSD |
| Current sourcing, Drive capability | 1200 Ω 950 Ω at 20 mA in HART mode |
| Current simulation, Range | 0 to 20 mA or 4 to 20 mA |
| Current simulation, Accuracy | 0.015% Rdg + 2 LSD |
| Current simulation, Ext. loop voltage requirement | 12-30 V |
| Loop power | While measuring mA, 24 V |
| Voltage measurement, Range | 0 to 28 V DC |
| Voltage measurement, Resolution | 1 mV |
| Voltage measurement, Accuracy | 0.015% Rdg + 2 LSD |
| Operating temperature | -10°C to 55°C |
|---|---|
| Storage temperature | -30°C to 60°C |
| Humidity (without condensation) | 0 to 90% (0°C to 35°C); 0 to 70% (35°C to 55°C) |
| Temperature coefficient | -10°C to 18°C, 28°C to 55°C, ±.005% of range per °C |
| Safety | CSA C22.2 No. 1010.1:1992 |
|---|---|
| Maximum voltage | 30 V |
| Size | 152 x 70 x 41 mm |
|---|---|
| Weight | 224 g (340 g including holster) |
| Batteries | One 9 V battery Alkaline |
| Warranty | Three years |
| Shock and vibration | Per MIL-T-28800 for a Class 2 Instrument |
| Display current and % of span | Yes |
| Auto step, auto ramp | Yes |
| Span check | Yes |
| Contents | AC70A Alligator Clips; C10 protective yellow holster with Flex-Stand; Instruction sheet; TL75 safety-designed test lead set; NIST- traceable calibration report and data |
|---|
Specifications are from the manufacturer's published information. Confirm critical values with us or the datasheet before ordering.
About the 707
The Fluke 707 combines push buttons and a quick-click knob so loop checks can be made quickly with one hand. It is a high-performance calibrator for calibration, repair and maintenance of current loops, with simultaneous mA and % readout on a large display.